Εμφάνιση απλής εγγραφής

dc.contributor.author Δαυίδ, Στράτος el
dc.contributor.author Γεωργίου, Μαρία el
dc.contributor.author Φυσικόπουλος, Ελευθέριος el
dc.contributor.author Λούντος, Γεώργιος el
dc.contributor.author Belcari, N. en
dc.date.accessioned 2015-06-05T16:12:55Z
dc.date.available 2015-06-05T16:12:55Z
dc.date.issued 2015-06-05
dc.identifier.uri http://hdl.handle.net/11400/15148
dc.rights Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες *
dc.rights.uri http://creativecommons.org/licenses/by-nc-nd/3.0/us/ *
dc.source http://iopscience.iop.org/ en
dc.subject Αξονική τομογραφία
dc.subject Silicon photomultipliers
dc.subject Ανιχνευτές σπινθηρισμών
dc.subject Φωτοπολλαπλασιαστές πυριτίου
dc.title Imaging performance of silicon photomultipliers coupled to BGO and CsI en
heal.type journalArticle
heal.secondaryTitle na arrays en
heal.classification Medicine
heal.classification Medical physics
heal.classification Ιατρική
heal.classification Ιατρική φυσική
heal.classificationURI http://id.loc.gov/authorities/subjects/sh00006614
heal.classificationURI http://id.loc.gov/authorities/subjects/sh85083001
heal.classificationURI **N/A**-Ιατρική
heal.classificationURI **N/A**-Ιατρική φυσική
heal.identifier.secondary DOI: 10.1088/1748-0221/8/12/P12008
heal.language en
heal.access campus
heal.publicationDate 2013
heal.bibliographicCitation David, S., Georgiou, M., Fysikopoulos, E., Belcari, N. and Loudos, G. (2013) Imaging performance of silicon photomultipliers coupled to BGO and CsI:Na arrays. "Journal of Instrumentation", 8 (12) en
heal.abstract The aim of this study is to investigate the imaging performance of a silicon photomultiplier array (ArraySL-4) photodetector for possible PET and potentially SPECT applications using BGO and CsI(Na) pixellated scintillators. Our main objectives are: i) the comparison of the ArraySL-4 to the older version SensL’s SPMArray4 photo detector in terms of energy resolution and peak to valley ratio of a row profile in the flood image and ii) the study of the effect of different coupling schemes using ultra transmitting glass windows of various thicknesses. We acquired raw images from two pixellated scintillators (BGO with 2 2 5mm3 and CsI:Na with 1 1 5mm3 pixel sizes) irradiated with 511 keV and 1274.5 keV g-rays from a 22Na source. The SiPM array detector allowed the clear visualization of the discrete 2 2mm2 pixellated BGO and 1 1mm2 CsI:Na scintillator elements at room temperature (no cooling). The energy resolution of the new SensL ArraySL-4 detector for the 2 2 5mm3 BGO pixellated scintillator array is improved for rather 6 percentage points (energy resolution improvement equal to 22%) and the peak to valley ratio is measured higher for both scintillator arrays (for BGO 68% (1:7 ) and for CsI:Na 154% (2:5 )) compared with SPMArray4. The clear identification of the 1 1mm2 CsI:Na scintillator elements provides evidence that the combination of those SiPMs with even smaller arrays can be used as an efficient imaging detector module. Optical coupling significantly improves image uniformity, while the use of BK7 ultra transmitting glass window with 1.35mm thickness provided the best measure energy resolution equal to 21.5%. en
heal.journalName Journal of Instrumentation en
heal.journalType peer-reviewed
heal.fullTextAvailability true


Αρχεία σε αυτό το τεκμήριο

  • Όνομα: 1748-0221_8_12_P12008.pdf
    Μέγεθος: 1022.Kb
    Μορφότυπο: PDF

Οι παρακάτω άδειες σχετίζονται με αυτό το τεκμήριο:

Εμφάνιση απλής εγγραφής

Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες Εκτός από όπου ορίζεται κάτι διαφορετικό, αυτή η άδεια περιγράφεται ως Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες