Εμφάνιση απλής εγγραφής

dc.contributor.author Σπύρου, Σπυρίδων Β. el
dc.contributor.author Μακρής, Δημήτριος el
dc.contributor.author Λούντος, Γεώργιος el
dc.date.accessioned 2015-06-07T10:14:17Z
dc.date.available 2015-06-07T10:14:17Z
dc.date.issued 2015-06-07
dc.identifier.uri http://hdl.handle.net/11400/15444
dc.rights Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες *
dc.rights.uri http://creativecommons.org/licenses/by-nc-nd/3.0/us/ *
dc.source http://www.sciencedirect.com/science/article/pii/S1120179715001180 en
dc.subject Monte Carlo
dc.subject Gold nanoparticles
dc.subject Μόντε Κάρλο
dc.subject Χρυσά νανοσωματίδια
dc.title Does the setup of Monte Carlo simulations influence the calculated properties and effect of gold nanoparticles in radiation therapy? en
heal.type report
heal.classification Medicine
heal.classification Biomedical engineering
heal.classification Ιατρική
heal.classification Βιοϊατρική τεχνολογία
heal.classificationURI http://id.loc.gov/authorities/subjects/sh00006614
heal.classificationURI http://id.loc.gov/authorities/subjects/sh85014237
heal.classificationURI **N/A**-Ιατρική
heal.classificationURI **N/A**-Βιοϊατρική τεχνολογία
heal.identifier.secondary doi:10.1016/j.ejmp.2015.05.008
heal.language en
heal.access campus
heal.recordProvider Τ.Ε.Ι. Αθήνας. Σχολή Τεχνολογικών Εφαρμογών. Τμήμα Μηχανικών Βιοϊατρικής Τεχνολογίας Τ.Ε. el
heal.publicationDate 2015
heal.abstract Purpose To investigate whether the dose-scoring process of Monte Carlo (MC) simulations of Gold nanoparticles (GNPs) in radiation therapy affects the results. Methods The GATE MC toolkit was used to simulate the irradiation of a water phantom containing a single solid or hollow GNP with 250 kVp and 6 MV photons. The dose was scored in 20 nm × 20 nm × 50 μm, 100 nm × 100 nm × 50 μm and 200 nm × 200 nm × 50 μm volumes using dose-scoring voxels of size 1 nm × 1 nm × 50 μm, 10 nm × 10 nm × 50 μm, 50 nm × 50 nm × 50 μm and 100 nm × 100 nm × 50 μm Εxcess dose depth-dose (EDDD) curves and lateral beam profiles were used to compare the dose-scoring voxels. Results In a given volume, neither the EDDD curves nor the lateral beam profiles are affected by the size of the dose-scoring voxels, subject to noise and uncertainty. Certain features of the EDDD curves are clearly seen in larger volumes, but hidden within the uncertainty and noise levels in smaller volumes. For the lateral beam profiles, it is the larger volumes that result in misleading results and the smaller ones that give the expected results. However, the limited statistics result in asymmetries and skewness in the profiles. Conclusion For a given volume, the dose curves are not affected by the size of the dose-scoring voxels. However, the voxel size may hide or reveal the finer structure of the dose curves and/or may result in misleading curves. en
heal.publisher Elsevier B.V. en
heal.fullTextAvailability true


Αρχεία σε αυτό το τεκμήριο

  • Όνομα: 1-s2.0-S1120179715001180-main.pdf
    Μέγεθος: 425.4Kb
    Μορφότυπο: PDF

Οι παρακάτω άδειες σχετίζονται με αυτό το τεκμήριο:

Εμφάνιση απλής εγγραφής

Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες Εκτός από όπου ορίζεται κάτι διαφορετικό, αυτή η άδεια περιγράφεται ως Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες