Εμφάνιση απλής εγγραφής

dc.contributor.author Ασβεστάς, Παντελής Α. el
dc.contributor.author Δελήμπασης, Κωνσταντίνος Κ. el
dc.contributor.author Μουραβλιάνσκυ, Νικόλαος Α. el
dc.contributor.author Ματσόπουλος, Γεώργιος Κ. el
dc.date.accessioned 2015-01-13T16:14:13Z
dc.date.available 2015-01-13T16:14:13Z
dc.date.issued 2015-01-13
dc.identifier.uri http://hdl.handle.net/11400/3942
dc.rights Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες *
dc.rights.uri http://creativecommons.org/licenses/by-nc-nd/3.0/us/ *
dc.source http://www.hindawi.com/ en
dc.subject Software registration system
dc.subject Radiotherapy
dc.subject Λογισμικό σύστημα καταχώρησης
dc.subject Ραδιοθεραπεία
dc.title Esterr-pro: a setup verification software system using electronic portal imaging en
heal.type journalArticle
heal.classification Medicine
heal.classification Medical technology
heal.classification Ιατρική
heal.classification Ιατρικά όργανα και εξοπλισμός
heal.classificationURI http://id.loc.gov/authorities/subjects/sh00006614
heal.classificationURI http://skos.um.es/unescothes/C02465
heal.classificationURI **N/A**-Ιατρική
heal.classificationURI **N/A**-Ιατρικά όργανα και εξοπλισμός
heal.identifier.secondary doi: 10.1155/2007/61523
heal.language en
heal.access free
heal.recordProvider Τ.Ε.Ι. Αθήνας. Σχολή Τεχνολογικών Εφαρμογών. Τμήμα Μηχανικών Βιοϊατρικής Τεχνολογίας Τ.Ε. el
heal.publicationDate 2007
heal.bibliographicCitation Asvestas, P., Delibasis, K., Mouravliansky, N. and Matsopoulos, G. (2007). Esterr-pro: a setup verification software system using electronic portal imaging. International Journal of Biomedical Imaging. vol. 2007. pp. 1-11. Available from: http://www.ncbi.nlm.nih.gov [Accessed 29/11/2006] en
heal.abstract The purpose of the paper is to present and evaluate the performance of a new software-based registration system for patient setup verification, during radiotherapy, using electronic portal images. The estimation of setup errors, using the proposed system, can be accomplished by means of two alternate registration methods. (a) The portal image of the current fraction of the treatment is registered directly with the reference image (digitally reconstructed radiograph (DRR) or simulator image) using a modified manual technique. (b) The portal image of the current fraction of the treatment is registered with the portal image of the first fraction of the treatment (reference portal image) by applying a nearly automated technique based on self-organizing maps, whereas the reference portal has already been registered with a DRR or a simulator image. The proposed system was tested on phantom data and on data from six patients. The root mean square error (RMSE) of the setup estimates was 0.8 ± 0.3 (mean value ± standard deviation) for the phantom data and 0.3 ± 0.3 for the patient data, respectively, by applying the two methodologies. Furthermore, statistical analysis by means of the Wilcoxon nonparametric signed test showed that the results that were obtained by the two methods did not differ significantly (P value > 0.05). en
heal.publisher Hindawi Publishing Corporation en
heal.journalName International Journal of Biomedical Imaging en
heal.journalType peer-reviewed
heal.fullTextAvailability true


Αρχεία σε αυτό το τεκμήριο

  • Όνομα: ESTERR-PRO A setup verification ...
    Μέγεθος: 2.283Mb
    Μορφότυπο: PDF

Οι παρακάτω άδειες σχετίζονται με αυτό το τεκμήριο:

Εμφάνιση απλής εγγραφής