Όνομα Περιοδικού:e-Journal of Science & Technology e-Περιοδικό Επιστήμης & Τεχνολογίας
Fundamental imaging performance in terms of Modulation Transfer Function (MTF) was
investigated for a high resolution CMOS based imaging sensor and a commercial Computed
Radiography (CR) unit. The CMOS device consists of a 33.91 mg/cm2 Gd2O2S:Tb scintillating screen,
placed in direct contact with a CMOS photodiode array. The CMOS photodiode array, featuring
1200×1600 pixels with a pixel pitch of 22.5
μm
, was used as an optical photon detector. The MTF was
measured using the slanted-edge method (measuring the Edge Spread Function–ESF), as well as by
using pin-hole phantoms (measuring the Point Spread Function–PSF). The experimental procedure
was performed under the representative radiation quality (RQA) settings, RQA 5 (70 kVp digital
radiography) and RQA M2 (28 kVp digital mammography) recommended by the International
Electrotechnical Commission Reports 62220-1 and 62220-1-2 respectively. It was found that the
detector response function was linear for the exposure ranges under investigation. MTF for the pinhole
method was found to increase by diminishing pin-hole size up to the detector pitch (22.5
μm
).
Additionally, our results showed that for the same RQA quality, MTF was comparable in the whole
spatial frequency range by both ESF and PSF (50
μm
) methods. MTF of the CMOS imaging sensor
was found better compared with the CR unit as well as compared with previously published data for other CCD and CMOS sensor.
Πηγή:[χ.ό.]
Θέμα:Medicine, Medical technology, Ιατρική, Ιατρική τεχνολογία