dc.contributor.author | Μιχαήλ, Χρήστος Μ. | el |
dc.contributor.author | Βαλαής, Ιωάννης Γ. | el |
dc.contributor.author | Τουτουντζής, Ανδριανός Ε. | el |
dc.contributor.author | Σεφέρης, Ιωάννης Ε. | el |
dc.contributor.author | Γεωργούσης, Μιχάλης | el |
dc.date.accessioned | 2015-01-22T11:06:02Z | |
dc.date.available | 2015-01-22T11:06:02Z | |
dc.date.issued | 2015-01-22 | |
dc.identifier.uri | http://hdl.handle.net/11400/4529 | |
dc.rights | Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες | * |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/3.0/us/ | * |
dc.source | http://link.springer.com/ | en |
dc.subject | Luminescence efficiency | |
dc.subject | Powder phosphors | |
dc.subject | Αποδοτικότητα φωταύγειας | |
dc.subject | Φωσφορίζουσα σκόνη | |
dc.title | Efficiency of Gd2O2S: eu powder phosphor as X-ray to light converter under radiographic imaging conditions | en |
heal.type | bookChapter | |
heal.generalDescription | IFMBE Proceedings | en |
heal.classification | Medicine | |
heal.classification | Physics | |
heal.classification | Ιατρική | |
heal.classification | Φυσική | |
heal.classificationURI | http://id.loc.gov/authorities/subjects/sh00006614 | |
heal.classificationURI | http://skos.um.es/unescothes/C02994 | |
heal.classificationURI | **N/A**-Ιατρική | |
heal.classificationURI | **N/A**-Φυσική | |
heal.contributorName | Φούντος, Γεώργιος Π. | el |
heal.contributorName | Σαμαρτζής, Αναστάσιος | el |
heal.contributorName | Λιαπαρίνος, Παναγιώτης Φ. | el |
heal.contributorName | Κανδαράκης, Ιωάννης Σ. | el |
heal.contributorName | Παναγιωτάκης, Γεώργιος Σ. | el |
heal.identifier.secondary | DOI 10.1007/978-3-642-03879-2_222 | |
heal.language | en | |
heal.access | free | |
heal.recordProvider | Τ.Ε.Ι. Αθήνας. Σχολή Τεχνολογικών Εφαρμογών. Τμήμα Μηχανικών Βιοϊατρικής Τεχνολογίας Τ.Ε. | el |
heal.publicationDate | 2009 | |
heal.bibliographicCitation | Michail, C., Valais, I., Toutountzis, A., Seferis, I., Georgousis, M., et al. (2009). Efficiency of Gd2O2S: eu powder phosphor as X-ray to light converter under radiographic imaging conditions. World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany. 25(2). pp. 794-797. Springer Berlin Heidelberg. Available from: http://link.springer.com | en |
heal.abstract | The aim of the present study was to examine the Gd2O2S:Eu powder scintillator under general X-ray Radiographic imaging conditions. For this purpose five scintillator screens with coating thicknesses 33.1, 46.4, 63.1, 78.3 and 139.8 mg/cm2 were prepared in our laboratory by sedimentation of Gd2O2S:Eu powder. Light emission efficiency and optical emission spectra of the screens were measured under X-ray excitation using X-ray tube voltages (50-140 kVp) employed in X-ray radiography. Spectral compatibility with various optical photon detectors (photodiodes, photocathodes, charge coupled devices, films) were determined using emission spectrum data. In addition, parameters related to X-ray detection and energy absorption efficiency were calculated. Gd2O2S:Eu showed peak emission in the wavelength range 620-630 nm. The 139.8 mg/cm2 phosphor screen appeared with the maximum light emission efficiency. Due to its reddish emission spectrum, Gd2O2S:Eu showed excellent compatibility with the sensitivity of many currently used photodetectors and could be considered for application in X-ray imaging especially in various digital detectors. | en |
heal.publisher | Springer Berlin Heidelberg | el |
heal.fullTextAvailability | true | |
heal.bookName | World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany | en |
Οι παρακάτω άδειες σχετίζονται με αυτό το τεκμήριο: