Εμφάνιση απλής εγγραφής

dc.contributor.author Νταλές, Χριστόφορος el
dc.contributor.author Κυνατίδης, Νικόλαος el
dc.contributor.author Μιχαήλ, Χρήστος Μ. el
dc.contributor.author Σεφέρης, Ιωάννης Ε. el
dc.contributor.author Βαλαής, Ιωάννης Γ. el
dc.date.accessioned 2015-02-03T17:18:47Z
dc.date.available 2015-02-03T17:18:47Z
dc.date.issued 2015-02-03
dc.identifier.uri http://hdl.handle.net/11400/5583
dc.rights Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες *
dc.rights.uri http://creativecommons.org/licenses/by-nc-nd/3.0/us/ *
dc.source http://e-jst.teiath.gr/ en
dc.subject MTF
dc.subject Nuclear medicine
dc.subject Slanted-edge Method
dc.subject Πυρηνική ιατρική
dc.subject CMOS
dc.subject CR
dc.subject Image Quality
dc.subject Μέθοδος Slanted-edge
dc.subject Pin-hole Phantoms
dc.title Image quality assessment in CMOS and cr medical imaging systems en
heal.type journalArticle
heal.generalDescription Special issue: Workshop on Bio-Medical Instrumentation and related Engineering And Physical Sciences, Technological Educational Institute of Athens, 6 July 2012 en
heal.classification Medicine
heal.classification Medical physics
heal.classification Ιατρική
heal.classification Ιατρική φυσική
heal.classificationURI http://id.loc.gov/authorities/subjects/sh00006614
heal.classificationURI http://id.loc.gov/authorities/subjects/sh85083001
heal.classificationURI **N/A**-Ιατρική
heal.classificationURI **N/A**-Ιατρική φυσική
heal.contributorName Καλύβας, Νεκτάριος Ι. el
heal.contributorName Φούντος, Γεώργιος Π. el
heal.contributorName Κανδαράκης, Ιωάννης Σ. el
heal.language en
heal.access free
heal.publicationDate 2012
heal.bibliographicCitation Ntales, C., Kynatidis, N., Michail, C.M., Seferis, I.E., Valais, I.G., et al. (2012). Image quality assessment in CMOS and cr medical imaging systems. "e-Journal of Science & Technology". [Online] 7(3): 7-13. Available from: http://e-jst.teiath.gr/ en
heal.abstract Fundamental imaging performance in terms of Modulation Transfer Function (MTF) was investigated for a high resolution CMOS based imaging sensor and a commercial Computed Radiography (CR) unit. The CMOS device consists of a 33.91 mg/cm2 Gd2O2S:Tb scintillating screen, placed in direct contact with a CMOS photodiode array. The CMOS photodiode array, featuring 1200×1600 pixels with a pixel pitch of 22.5 μm , was used as an optical photon detector. The MTF was measured using the slanted-edge method (measuring the Edge Spread Function–ESF), as well as by using pin-hole phantoms (measuring the Point Spread Function–PSF). The experimental procedure was performed under the representative radiation quality (RQA) settings, RQA 5 (70 kVp digital radiography) and RQA M2 (28 kVp digital mammography) recommended by the International Electrotechnical Commission Reports 62220-1 and 62220-1-2 respectively. It was found that the detector response function was linear for the exposure ranges under investigation. MTF for the pinhole method was found to increase by diminishing pin-hole size up to the detector pitch (22.5 μm ). Additionally, our results showed that for the same RQA quality, MTF was comparable in the whole spatial frequency range by both ESF and PSF (50 μm ) methods. MTF of the CMOS imaging sensor was found better compared with the CR unit as well as compared with previously published data for other CCD and CMOS sensor. en
heal.publisher Νερατζής, Ηλίας el
heal.publisher Σιανούδης, Ιωάννης el
heal.publisher Βαλαής, Ιωάννης Γ. el
heal.publisher Φούντος, Γεώργιος Π. el
heal.journalName e-Journal of Science & Technology en
heal.journalName e-Περιοδικό Επιστήμης & Τεχνολογίας el
heal.journalType peer-reviewed
heal.fullTextAvailability true


Αρχεία σε αυτό το τεκμήριο

Οι παρακάτω άδειες σχετίζονται με αυτό το τεκμήριο:

Εμφάνιση απλής εγγραφής

Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες Εκτός από όπου ορίζεται κάτι διαφορετικό, αυτή η άδεια περιγράφεται ως Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες