Εμφάνιση απλής εγγραφής

dc.contributor.author Δαυίδ, Στράτος Λ. el
dc.contributor.author Γεωργίου, Μαρία el
dc.contributor.author Φυσικόπουλος, Ελευθέριος el
dc.contributor.author Λούντος, Γεώργιος el
dc.date.accessioned 2015-02-03T17:26:57Z
dc.date.available 2015-02-03T17:26:57Z
dc.date.issued 2015-02-03
dc.identifier.uri http://hdl.handle.net/11400/5584
dc.rights Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες *
dc.rights.uri http://creativecommons.org/licenses/by-nc-nd/3.0/us/ *
dc.source http://e-jst.teiath.gr/ el
dc.subject Nuclear medicine
dc.subject Πυρηνική ιατρική
dc.subject Nuclear Imaging
dc.subject Πυρηνική απεικόνιση
dc.subject LuAG:Pr
dc.subject PSPMTs
dc.subject Scintillator Crystals
dc.subject Κρύσταλλοι σπινθηριστή
dc.title Initial results for the development of a small field of view gamma-ray imager using a LuAG:Pr scintillator en
heal.type journalArticle
heal.generalDescription Special issue: Workshop on Bio-Medical Instrumentation and related Engineering And Physical Sciences, Technological Educational Institute of Athens, 6 July 2012 en
heal.classification Medicine
heal.classification Medical physics
heal.classification Ιατρική
heal.classification Ιατρική φυσική
heal.classificationURI http://id.loc.gov/authorities/subjects/sh00006614
heal.classificationURI http://id.loc.gov/authorities/subjects/sh85083001
heal.classificationURI **N/A**-Ιατρική
heal.classificationURI **N/A**-Ιατρική φυσική
heal.keywordURI http://id.loc.gov/authorities/subjects/sh85093011
heal.language en
heal.access free
heal.publicationDate 2012
heal.bibliographicCitation David S.L., Georgiou, M., Fysikopoulos, E. and Loudos, G. (2012). Initial results for the development of a small field of view gamma-ray imager using a LuAG:Pr scintillator. "e-Journal of Science & Technology". [Online] 7(3): 15-21. Available from: http://e-jst.teiath.gr/ en
heal.abstract In this study we present the initial results for the construction and the evaluation of a small field of view gamma ray detector based on a R8900U-00-C12 position sensitive photomultiplier tube (PSPMT) coupled to a 2x2 LuAG:Pr scintillator array with 5x5x5mm3 crystal size elements. We have used a resistor network readout circuit, reducing the 6X and 6Y photomultiplier anode signals to 4 position signals (Xa, Xb, Yc and Yd). A Field Programmable Gate Array (FPGA) Spartan 6 LX16 was used for triggering and signal processing of the signal pulses acquired using a free running sampling technique. We have obtained raw images and energy histograms, under various gamma ray energies irradiation. Energy resolution and linearity of the detector response under various gamma ray energies ranging from 81 to 1332.5 keV were measured. The system presented an excellent performance in terms of energy resolution, measured 9.3%@511KeV and 8.5%@662keV. en
heal.publisher Νερατζής, Ηλίας el
heal.publisher Σιανούδης, Ιωάννης el
heal.publisher Βαλαής, Ιωάννης Γ. el
heal.publisher Φούντος, Γεώργιος Π. el
heal.journalName e-Journal of Science & Technology en
heal.journalName e-Περιοδικό Επιστήμης & Τεχνολογίας el
heal.journalType peer-reviewed
heal.fullTextAvailability true


Αρχεία σε αυτό το τεκμήριο