Εμφάνιση απλής εγγραφής

dc.contributor.author Σεφέρης, Ιωάννης Ε. el
dc.contributor.author Μιχαήλ, Χρήστος Μ. el
dc.contributor.author Βαλαής, Ιωάννης Γ. el
dc.contributor.author Φούντος, Γεώργιος Π. el
dc.contributor.author Καλύβας, Νεκτάριος Ι. el
dc.date.accessioned 2015-02-03T17:45:18Z
dc.date.available 2015-02-03T17:45:18Z
dc.date.issued 2015-02-03
dc.identifier.uri http://hdl.handle.net/11400/5587
dc.rights Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες *
dc.rights.uri http://creativecommons.org/licenses/by-nc-nd/3.0/us/ *
dc.source http://e-jst.teiath.gr/ en
dc.subject Nuclear medicine
dc.subject Πυρηνική ιατρική
dc.subject MTF
dc.subject Image degradation
dc.subject Scintillating screen
dc.subject Σπινθηρίζουσα οθόνη
dc.subject Υποβάθμιση εικόνας
dc.title X-ray image degradation passing through thin glass substrate en
heal.type journalArticle
heal.generalDescription Special issue: Workshop on Bio-Medical Instrumentation and related Engineering And Physical Sciences, Technological Educational Institute of Athens, 6 July 2012 en
heal.classification Medicine
heal.classification Medical physics
heal.classification Ιατρική
heal.classification Ιατρική φυσική
heal.classificationURI http://id.loc.gov/authorities/subjects/sh00006614
heal.classificationURI http://id.loc.gov/authorities/subjects/sh85083001
heal.classificationURI **N/A**-Ιατρική
heal.classificationURI **N/A**-Ιατρική φυσική
heal.keywordURI http://id.loc.gov/authorities/subjects/sh85093011
heal.contributorName Στρωματιά, Φωτεινή el
heal.contributorName Κανδαράκης, Ιωάννης Σ. el
heal.contributorName Παναγιωτάκης, Γεώργιος Σ. el
heal.language en
heal.access free
heal.publicationDate 2012
heal.bibliographicCitation Seferis, I.E., Michail, C.M., Valais, I.G., Fountos, G.P., Kalyvas, N.I., et al. (2012). X-ray image degradation passing through thin glass substrate. "e-Journal of Science & Technology". [Online] 7(3): 29-31. Available from: http://e-jst.teiath.gr/ en
heal.abstract The degradation of X-ray image passing through thin glass substrate was experimentally investigated in term of modulation transfer function (MTF). The screen was prepared by sedimentation of the powder phosphor (Gd2O2S:Eu) on thin glass substrate with thickness 140 m (Waldemar Knittel- GmbH). The experiment was carried out under X-ray mammographic conditions. The beam quality was 28 kV, 60 mAs, with W/Rh target/filter combination. The screen was coupled to an optical readout device including a CMOS remote RadEye HR photodiode pixel array. Results showed that there is a 40% degradation of the MTF on average, in the whole spatial frequency range, due to the glass substrate. The glass substrate has a low impact on resolution of the image at low frequencies (12% on average, up to 2.7 cy/mm) and a high impact at medium and high frequencies (70% on average). en
heal.publisher Νερατζής, Ηλίας el
heal.publisher Σιανούδης, Ιωάννης el
heal.publisher Βαλαής, Ιωάννης Γ. el
heal.publisher Φούντος, Γεώργιος Π. el
heal.journalName e-Journal of Science & Technology en
heal.journalName e-Περιοδικό Επιστήμης & Τεχνολογίας el
heal.journalType peer-reviewed
heal.fullTextAvailability true


Αρχεία σε αυτό το τεκμήριο

Οι παρακάτω άδειες σχετίζονται με αυτό το τεκμήριο:

Εμφάνιση απλής εγγραφής

Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες Εκτός από όπου ορίζεται κάτι διαφορετικό, αυτή η άδεια περιγράφεται ως Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες