dc.contributor.author | Κάβουρας, Διονύσης Α. | el |
dc.contributor.author | Κανδαράκης, Ιωάννης Σ. | el |
dc.contributor.author | Παναγιωτάκης, Γεώργιος Σ. | el |
dc.contributor.author | Κανελλόπουλος, Ε. | el |
dc.contributor.author | Τριάντης, Δήμος Α. | el |
dc.date.accessioned | 2015-04-14T18:08:19Z | |
dc.date.available | 2015-04-14T18:08:19Z | |
dc.date.issued | 2015-04-14 | |
dc.identifier.uri | http://hdl.handle.net/11400/8373 | |
dc.rights | Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες | * |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/3.0/us/ | * |
dc.source | http://www.elsevier.com | en |
dc.subject | Modulation transfer function (MTF) | |
dc.subject | Test screens | |
dc.subject | Sedimentation of the phosphor powder | |
dc.subject | Δοκιμή οθονών | |
dc.subject | Καθίζηση της σκόνης φωσφόρου | |
dc.subject | Συνάρτηση μεταφοράς διαμόρφωσης | |
dc.title | An investigation of the imaging characteristics of the Y2O2S:Eu3+ phosphor for application in X-ray detectors of digital mammography | en |
heal.type | journalArticle | |
heal.classification | Medicine | |
heal.classification | Medical physics | |
heal.classification | Ιατρική | |
heal.classification | Ιατρική φυσική | |
heal.classificationURI | http://id.loc.gov/authorities/subjects/sh00006614 | |
heal.classificationURI | http://id.loc.gov/authorities/subjects/sh85083001 | |
heal.classificationURI | **N/A**-Ιατρική | |
heal.classificationURI | **N/A**-Ιατρική φυσική | |
heal.identifier.secondary | doi:10.1016/S0969-8043(97)10108-7 | |
heal.language | en | |
heal.access | free | |
heal.publicationDate | 1998-05-11 | |
heal.bibliographicCitation | CAVOURAS, D.A., KANDARAKIS, I.S., PANAYIOTAKIS, G.S., KANELLOPOULOS, E., TRIANTIS, D.A., ET AL. (1998). An investigation of the imaging characteristics of the Y2O2S:Eu3+ phosphor for application in X-ray detectors of digital mammography . Applied Radiation and Isotopes. [Online] 49 (8). p.931-937. Available from: http://www.sciencedirect.com/[Accessed 22/11/1998] | en |
heal.abstract | Y2O2S:Eu laboratory prepared screens were evaluated as mammographic image receptors and were compared to similarly prepared screens of Gd2O2S:Tb and Y2O2S:Tb phosphor materials, often used in X-ray imaging detectors. The evaluation was performed by determining the Modulation Transfer Function (MTF) and the spatial frequency dependent Detective Quantum Efficiency (DQE). Y2O2S:Eu exhibited higher DQE values at low frequencies and given its good spectral matching with digital optical detectors, it may be appropriate for use in X-ray digital mammography. | en |
heal.publisher | Elsevier | en |
heal.journalName | Applied Radiation and Isotopes | en |
heal.journalType | peer-reviewed | |
heal.fullTextAvailability | true |