Εμφάνιση απλής εγγραφής

dc.contributor.author Κάβουρας, Διονύσης Α. el
dc.contributor.author Κανδαράκης, Ιωάννης Σ. el
dc.contributor.author Παναγιωτάκης, Γεώργιος Σ. el
dc.contributor.author Κανελλόπουλος, Ε. el
dc.contributor.author Τριάντης, Δήμος Α. el
dc.date.accessioned 2015-05-04T10:22:41Z
dc.date.available 2015-05-04T10:22:41Z
dc.date.issued 2015-05-04
dc.identifier.uri http://hdl.handle.net/11400/9634
dc.rights Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες *
dc.rights.uri http://creativecommons.org/licenses/by-nc-nd/3.0/us/ *
dc.source http://www.sciencedirect.com/science/article/pii/S0969804397101087 en
dc.subject Mammographic image receptors
dc.subject Phosphor
dc.subject Υποδοχείς μαστογραφικών εικόνων
dc.subject Φώσφορος
dc.title An investigation of the imaging characteristics of the Y2O2S:Eu phosphor for application in x-ray detectors of digital mammography en
heal.type journalArticle
heal.classification Medicine
heal.classification Physics
heal.classification Ιατρική
heal.classification Φυσική
heal.classificationURI http://id.loc.gov/authorities/subjects/sh00006614
heal.classificationURI http://skos.um.es/unescothes/C02994
heal.classificationURI **N/A**-Ιατρική
heal.classificationURI **N/A**-Φυσική
heal.contributorName Νομικός, Κωνσταντίνος Δ. el
heal.identifier.secondary doi:10.1016/S0969-8043(97)10108-7
heal.language en
heal.access campus
heal.recordProvider Τ.Ε.Ι. Αθήνας. Σχολή Τεχνολογικών Εφαρμογών. Τμήμα Μηχανικών Βιοϊατρικής Τεχνολογίας Τ.Ε. el
heal.publicationDate 1998
heal.bibliographicCitation Cavouras, D., Kandarakis, I., Panayiotakis, G., Kanellopoulos, E., Triantis, D., et al. (May 1998). An investigation of the imaging characteristics of the Y2O2S:Eu phosphor for application in x-ray detectors of digital mammography. Applied Radiation and Isotopes. 49(8). pp. 931-937. Available from: http://www.sciencedirect.com/science/article/pii/S0969804397101087 [Accessed 22/11/1998] en
heal.abstract Y2O2S:Eu laboratory prepared screens were evaluated as mammographic image receptors and were compared to similarly prepared screens of Gd2O2S:Tb and Y2O2S:Tb phosphor materials, often used in X-ray imaging detectors. The evaluation was performed by determining the Modulation Transfer Function (MTF) and the spatial frequency dependent Detective Quantum Efficiency (DQE). Y2O2S:Eu exhibited higher DQE values at low frequencies and given its good spectral matching with digital optical detectors, it may be appropriate for use in X-ray digital mammography. en
heal.publisher Elsevier Science Ltd en
heal.journalName Applied Radiation and Isotopes en
heal.journalType peer-reviewed
heal.fullTextAvailability true


Αρχεία σε αυτό το τεκμήριο

  • Όνομα: An investigation of the imaging ...
    Μέγεθος: 538.4Kb
    Μορφότυπο: PDF

Οι παρακάτω άδειες σχετίζονται με αυτό το τεκμήριο:

Εμφάνιση απλής εγγραφής

Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες Εκτός από όπου ορίζεται κάτι διαφορετικό, αυτή η άδεια περιγράφεται ως Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες