Εμφάνιση απλής εγγραφής

dc.contributor.author Κανδαράκης, Ιωάννης Σ. el
dc.contributor.author Κάβουρας, Διονύσης Α. el
dc.contributor.author Κανελλόπουλος, Ε. el
dc.contributor.author Νομικός, Κωνσταντίνος Δ. el
dc.contributor.author Παναγιωτάκης, Γεώργιος Σ. el
dc.date.accessioned 2015-05-04T12:30:07Z
dc.date.available 2015-05-04T12:30:07Z
dc.date.issued 2015-05-04
dc.identifier.uri http://hdl.handle.net/11400/9655
dc.rights Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες *
dc.rights.uri http://creativecommons.org/licenses/by-nc-nd/3.0/us/ *
dc.source http://link.springer.com/article/10.1007%2FBF02513261 en
dc.subject Scintillators
dc.subject Information capacity
dc.subject Σπινθηριστές
dc.subject Πληροφοριακή ικανότητα
dc.title A method for determining the information capacity of x-ray scintillator detectors by means of luminesence and modulation transfer function measurements en
heal.type journalArticle
heal.classification Medicine
heal.classification Physics
heal.classification Ιατρική
heal.classification Φυσική
heal.classificationURI http://id.loc.gov/authorities/subjects/sh00006614
heal.classificationURI http://skos.um.es/unescothes/C02994
heal.classificationURI **N/A**-Ιατρική
heal.classificationURI **N/A**-Φυσική
heal.keywordURI http://id.loc.gov/authorities/subjects/sh85118713
heal.identifier.secondary DOI 10.1007/BF02513261
heal.language en
heal.access campus
heal.recordProvider Τ.Ε.Ι. Αθήνας. Σχολή Τεχνολογικών Εφαρμογών. Τμήμα Μηχανικών Βιοϊατρικής Τεχνολογίας Τ.Ε. el
heal.publicationDate 1999
heal.bibliographicCitation Kandarakis, I., Cavouras, D., Kanellopoulos, E., Nomicos, C. and Panayiotakis, G. (January 1999). A method for determining the information capacity of x-ray scintillator detectors by means of luminesence and modulation transfer function measurements. Medical & Biological Engineering & Computing. 37(1). pp. 25-30. Kluwer Academic Publishers: 1999. Available from: http://link.springer.com/article/10.1007%2FBF02513261 en
heal.abstract A method to determine the information capacity of x-ray phosphor screens used in the detectors of medical imaging systems is described. Information capacity was determined via x-ray luminescence efficiency (XLE), modulation transfer function (MTF) and emission spectrum measurements. The method was applied to laboratory prepared screens from commonly employed phosphor materials. The screen coating weight varied from 50 mg cm−2 to 140 mg cm−2. Results indicated that information capacity decreased with screen coating thickness but also depended on intrinsic phosphor properties (density, effective atomic number, intrinsic conversion efficiency, light wavelength). The Gd2O2S:Tb phosphor, exhibiting high density and effective atomic number, was found to be superior to La2O2S:Tb and Y2O2S:Tb. en
heal.publisher Kluwer Academic Publishers en
heal.journalName Medical & Biological Engineering & Computing en
heal.journalType peer-reviewed
heal.fullTextAvailability true


Αρχεία σε αυτό το τεκμήριο

  • Όνομα: A method for determining the ...
    Μέγεθος: 546.3Kb
    Μορφότυπο: PDF

Οι παρακάτω άδειες σχετίζονται με αυτό το τεκμήριο:

Εμφάνιση απλής εγγραφής

Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες Εκτός από όπου ορίζεται κάτι διαφορετικό, αυτή η άδεια περιγράφεται ως Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες