Εμφάνιση απλής εγγραφής

dc.contributor.author Φυσικόπουλος, Ελευθέριος el
dc.contributor.author Γεωργίου, Μαρία el
dc.contributor.author Λούντος, Γεώργιος el
dc.contributor.author Ματσόπουλος, Γεώργιος el
dc.date.accessioned 2015-06-05T16:42:40Z
dc.date.available 2015-06-05T16:42:40Z
dc.date.issued 2015-06-05
dc.identifier.uri http://hdl.handle.net/11400/15156
dc.rights Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες *
dc.rights.uri http://creativecommons.org/licenses/by-nc-nd/3.0/us/ *
dc.source http://iopscience.iop.org en
dc.subject Digital electronic circuits
dc.subject Digital signal processing
dc.subject Performance parameters
dc.subject Παράμετροι απόδοσης
dc.subject Ψηφιακά ηλεκτρονικά κυκλώματα
dc.subject Ψηφιακή επεξεργασία σήματος
dc.title Low cost FPGA based data acquisition system for a gamma imaging probe en
heal.type journalArticle
heal.classification Medicine
heal.classification Medical physics
heal.classification Ιατρική
heal.classification Ιατρική φυσική
heal.classificationURI http://id.loc.gov/authorities/subjects/sh00006614
heal.classificationURI http://id.loc.gov/authorities/subjects/sh85083001
heal.classificationURI **N/A**-Ιατρική
heal.classificationURI **N/A**-Ιατρική φυσική
heal.identifier.secondary DOI: 10.1088/1748-0221/8/11/T11004
heal.language en
heal.access free
heal.publicationDate 2013
heal.bibliographicCitation Fysikopoulos, E., Georgiou, M., Loudos, G. and Matsopoulos, G. (2013) Low cost FPGA based data acquisition system for a gamma imaging probe. "Journal of Instrumentation", 8 (11) en
heal.abstract We present the development of a low cost field programmable gate arrays (FPGA) based data acquisition system for a gamma imaging probe proposed for sentinel lymph node (SLN) mapping. Radioguided surgery using a gamma probe is an established practice and has been widely introduced in SLN biopsies. For such applications, imaging systems require compact readout electronics and flexibility. Embedded systems implemented in the FPGA technology offer new possibilities in data acquisition for nuclear medicine imagers. FPGAs are inexpensive compared to application specific integrated circuits (ASICs), usually used for the readout electronics of dedicated gamma cameras and their size is rather small. In this study, cost effective analog to digital converters (ADCs) were used and signal processing algorithms were implemented in the FPGA to extract the energy and position information. The analog front-end electronics were carefully designed taking into account the low sampling rate of the ADCs. The reference gamma probe has a small field of view (2.5 cm × 2.5 cm) and is based on the R8900U-00-C12 position sensitive photomultiplier tube (PSPMT) coupled to a pixellated CsI(Na) scintillator with 1 mm × 1 mm × 5 mm crystal element size. Measurements were carried out using a general purpose collimator and 99mTc sources emitted at 140 keV. Performance parameters for the imaging gamma probe were compared with those obtained when data were acquired using the standard NIM (Nuclear Instrumentation Modules) electronics and found to be in very good agreement, which demonstrates the efficiency of the proposed implementation. en
heal.journalName Journal of Instrumentation en
heal.journalType peer-reviewed
heal.fullTextAvailability true


Αρχεία σε αυτό το τεκμήριο

  • Όνομα: 1748-0221_8_11_T11004.pdf
    Μέγεθος: 1.028Mb
    Μορφότυπο: PDF

Οι παρακάτω άδειες σχετίζονται με αυτό το τεκμήριο:

Εμφάνιση απλής εγγραφής

Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες Εκτός από όπου ορίζεται κάτι διαφορετικό, αυτή η άδεια περιγράφεται ως Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες