The aim of the present study is to simulate a commercial contrast detail
phantom - (CDRAD 2.0 phantom) used in the evaluation of both digital and analog
radiographic units. This phantom is suitable for the estimation of low contrast
resolution. The software phantom was developed in the Matlab 7.01 platform. The
results of the software phantom were evaluated with the ones obtained on a digital
radiographic unit (DIAGNOST Philips Medical Systems) and showed a very good
agreement. The software phantom uses tabulated data for the X-ray spectra and
attenuation coefficients in order to simulate the CDRAD images as well as the
contrast details versus X-ray energy and phantom depth. This software phantom
provides an easy, fast and reliable method for the evaluation of contrast detail in
radiographic units (both digital and analog) with the use of a personal computer.