The performance of a new calibrator for pulse oximeters is tested with five pulse oximeters from
different manufacturers. The calibrator is based on time resolved transmission spectra of human
fingers. Finger spectra with different arterial oxygen saturation can be selected to simulate real
patients. The results obtained with this calibration device are compared with the results of
conventional calibration procedures with volunteers. Beside accuracy tests the suitability for
artifact simulation with the new device is discussed. The response of the five tested pulse
oximeters is in good agreement with the response of the pulse oximeters connected to real
patients. A test procedure for pulse oximeters similar to the conventional desaturation practice is
possible; some of the typical artifacts pulse oximetry has to cope with can be simulated easily.