dc.title A two-layered flow induced by peristaltic waves in a catheterized tube en
heal.type journalArticle
heal.classification Technology
heal.classification Mechanical engineering
heal.classification Τεχνολογία
heal.classification Μηχανολογία
heal.classificationURI http://id.loc.gov/authorities/subjects/sh85133147
heal.classificationURI http://skos.um.es/unescothes/C02447
heal.classificationURI **N/A**-Τεχνολογία
heal.classificationURI **N/A**-Μηχανολογία
heal.language en
heal.access free
heal.publicationDate 2012
heal.bibliographicCitation Medhavi, A. (2012). A two-layered flow induced by peristaltic waves in a catheterized tube. "e-Journal of Science & Technology". [Online] 7(5): 29-41. Available from: http://e-jst.teiath.gr/ en
heal.abstract The flow of a two-layered Newtonian fluid induced by peristaltic waves in a catheterized tube has been investigated. The expressions for the flow characteristicsthe flow rate, the pressure drop and the friction forces at the tube and catheter wall are derived. It is found that the pressure drop increases with the flow rate but decreases with the increasing peripheral layer thickness and a linear relationship between pressure and flow exists. The pressure drop increases with the catheter size (radius) and assumes a high asymptotic at the catheter size more that fifty percent of the tube size. The friction forces at the tube and catheter wall posses characteristics similar to that of the pressure drop with respect to any parameter. However, friction force at catheter wall assumes much smaller magnitude than the corresponding value at the tube wall. en
heal.publisher Νερατζής, Ηλίας el
heal.publisher Σιανούδης, Ιωάννης el
heal.journalName e-Journal of Science & Technology en
heal.journalName e-Περιοδικό Επιστήμης & Τεχνολογίας el
heal.journalType peer-reviewed
heal.fullTextAvailability true


Αρχεία σε αυτό το τεκμήριο

Οι παρακάτω άδειες σχετίζονται με αυτό το τεκμήριο:

Εμφάνιση απλής εγγραφής

Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες Εκτός από όπου ορίζεται κάτι διαφορετικό, αυτή η άδεια περιγράφεται ως Αναφορά Δημιουργού-Μη Εμπορική Χρήση-Όχι Παράγωγα Έργα 3.0 Ηνωμένες Πολιτείες