Όνομα Περιοδικού:International Journal of Biomedical Imaging
The purpose of the paper is to present
and evaluate the performance of a new softwarebased
registration system for patient set-up
verification, during radiotherapy, using electronic
portal images. The estimation of set-up errors, using
the proposed system, can be accomplished by means
of registering the portal image of the current
fraction of the treatment with the portal image of the
first image of the treatment (reference portal image)
by applying a nearly automated technique. The
proposed system was tested on data obtained by a
phantom as well as on data from four patients. The
phantom data included portal images that
corresponded to various positions of the phantom on
the treatment couch. For each patient, a set of thirty
portal images were used. The accuracy of the system
was (in average) better than 1 mm and the estimated
set-up error was very close to the expected one.